Skip to product information
1 of 3

K9400FU

Brand: Yokogawa
Order Number: K9400FU
Type: Gasket
Country:
Low stock: 15 left

Description

Caret Down

The Yokogawa K9400FU is a replacement gasket used in the air turbine assembly for the agitator fan of a process gas chromatograph. It is installed with the gasket-holding plate as part of the fan assembly, providing a sealing interface during reassembly and maintenance. The part is identified in Yokogawa's GC8000 maintenance documentation as a replacement component.

Technical Specifications

Parameter Specification
Model K9400FU
Brand Yokogawa
Product Type Gasket
Assembly Air Turbine for Agitator Fan
Equipment Process Gas Chromatograph
Function Sealing component for agitator fan assembly

Function and Application

The K9400FU gasket belongs to the agitator fan assembly associated with the isothermal oven. The assembly includes an air turbine, fan and gasket-holding plate. The gasket is fitted between the relevant assembly surfaces to maintain the intended sealing interface.

During maintenance, the gasket is removed together with its holding plate when accessing the air turbine. Yokogawa's maintenance procedure specifies reinstalling the gasket and holding plate when reassembling the fan components.

Maintenance and Replacement

The gasket should be inspected when the agitator fan or air turbine is serviced. If removed or damaged during maintenance, it should be replaced with the specified part to restore the assembly's sealing interface.

When ordering, verify the complete part number K9400FU against the equipment maintenance parts list. Do not confuse it with K9400FA, which is the separate air turbine component.

Frequently Asked Questions

Q: What is the function of the Yokogawa K9400FU?

A: It is a gasket for the air turbine assembly of the agitator fan. It provides the sealing interface required when the fan assembly is reassembled.

Q: Is K9400FU the same component as K9400FA?

A: No. K9400FU is the gasket, while K9400FA is the air turbine. They are separately identified replacement parts.

Q: When should the K9400FU gasket be replaced?

A: It should be inspected during air turbine or agitator fan maintenance and replaced if damaged or unsuitable for reuse. Follow the applicable equipment maintenance procedure.